Abstract

A simple, accurate and precise RP-HPLC method was developed for simultaneous estimation of Rosuvastatin and Bisoprolol fumarate in bulk and in formulation. Separation was achieved using and C18 column and Methanol: Phosphate buffer (pH 3.5) at a ratio of 45:55% V/V, flow rate was maintained at 1.0 ml. min and 245 nm was used as determination wavelength. System suitability parameters shows values within specified limits. Developed method was validated as per ICH guidelines and all the parameters were found to be with in the specified limits as per regulatory guidelines. Linearity was found to be 4-20 and 8-40 mcg/ml for Rosuvastatin and Bisoprolol fumarate respectively. Results of accuracy was found to be with in 98-102% which shows accuracy of the developed method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.