A new, reliable, and validated reverse phase-high-performance liquid chromatography (HPLC) method was developed to quantify the amount of allopurinol and lesinurad simultaneously in solid (tablet) dosage form. A clear chromatographic division was attained on inertsil ODS (4.6 x 250 mm, 5 mm) column, and a mixture of 0.1% trifluoroacetic acid and methanol in the ratio of 40:60 v/v was used as mobile phase. The rate of flow was set at 1 mL/min, and UV detection was achieved at λmax of 255 nm. Injection volume was set to 20 μL. The correlation coefficient of 0.999 was established, and the accurateness was found to be 100.69 and 100.49 for both the drugs, respectively. Therefore, the developed method was simple, specific, precise, and stable. Hence, the method can be employed to estimate the said drugs in other pharmaceutical formulations.