Abstract

A reverse-phase high-performance liquid chromatographic (RP-HPLC) method was developed to analyze the simultaneous estimation of doxorubicin and clotrimazole. The method was achieved by Nucleodur C18 column with dimension 250 × 4.6 mm (5 μm) using gradient elution. The mobile phase contained 0.2% formic acid (pH 3.2) and acetonitrile. The flow rate was kept at 1.0 mL/min and detection and quantitation of both drugs (doxorubicin and clotrimazole) were achieved using a photodiode array detector at 276 nm, which was the isosbestic point for both drugs. The proposed method was validated according to the current International Council for Harmonization of Technical Requirements of Pharmaceuticals for Human Use guidelines for specificity, linearity, accuracy, precision, and robustness. The developed method showed a linear response (R2 > 0.999), and was accurate (recoveries 97%-103%), precise (resolution ≤1.0%), sensitive, and specific. Thus, the developed RP-HPLC method for the simultaneous estimation of both drugs was successfully validated and can be utilized for the estimation of these drugs in the formulations being developed.