A sensitive liquid chromatographic–electrospray ionization mass spectrometric method was developed and validated for fast determination of telmisartan in human plasma. Plasma of 0.1 mL was deprotienated with methanol, centrifugation, evaporation to dryness and dissolving in mobile phase, samples were separated using a Hypersil-Keystone C18 reversed-phase column (150 mm × 2.1 mm i.d., 5 μm), together with a mobile phase containing of acetonitrile–10 mM ammonium acetate (42:58, v/v), 0.2% acetic acid and was isocratically eluted at a flow rate of 0.2 mL/min. Telmisartan and its internal standard, valsartan, were measured by electrospray ion source in positive selective ion monitoring mode. The method demonstrated linearity from 1 to 2000 ng/mL ( r = 0.9988). The limit of quantification for telmisartan in plasma was 1 ng/mL with good accuracy and precision. The mean sample extract recovery of the method were higher than 82 and 78% for telmisartan and internal standard (IS), respectively. The within-run and between-run precision ranged from 3.4 to 8.9% and 5.9 to 11.2% (relative standard deviation, R.S.D.), respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.