Abstract

A high-performance thin layer chromatographic method for determination of Teneligliptin was developed and validation as per ICH guidelines. HPTLC separation was performed on aluminum plates precoated with silica gel 60F254 and Methanol: Toluene: Triethylamine (1:3:1% v/v) as optimized mobile phase at detection wavelength of 245 nm. The retardation factor (Rf) value for Teneligliptin were 0.63 respectively. Accuracy for the marketed formulation teneza was found to be 98.31-100.51%.The percent relative standard deviation for repeatability and intermediate precision studies was found to be < 2%. The propose development HPTLC method can be applied for identification and quantitative determination of Teneligliptin

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.