UPLC-MS/MS analytical conditions for the analysis of aflatoxins in spices were optimized and validated in this study. Liquid-liquid partition-based protocols for the cleaning up of extracts using common organic solvents such as acetonitrile, hexane, and ethyl acetate were developed and validated. The developed liquid-liquid partition methods were compared with immuno-affinity column and QuEChERS clean-up methods for the UPLC-MS/MS analysis of aflatoxins in 8 spices. The reduction of lipophilic components using the partition with hexane is particularly useful in spices like red pepper that have higher levels of fatty acids, carotenoids, sterols, triterpenoids, etc. The subsequent partitioning with ethyl acetate considerably reduced the matrix interference from the polar components and increased the sensitivity. The cleaning up of spice extracts using liquid-liquid partition techniques resulted in limits of quantification (LOQ) of 2-5 µgL-1 in UPLC-MS/MS analysis. Trueness, repeatability, and reproducibility of the methods were in acceptable ranges. The accuracy of the developed methods was further verified by analyzing aflatoxins in naturally incurred samples of spices and comparing the results with those obtained from the immuno-affinity column cleanup-HPLC-FD method.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.