Abstract

A stereospecific HPLC method for the quantitation of CGP 49309 in samples of its corresponding enantiomer valsartan has been developed and validated. The enantiomeric separation was achieved on a 5 micron silica-bonded, alpha 1-acid glycoprotein column (Chiral AGP) with a phosphate buffer, pH 7, containing 2% (v/v) 2-propanol as a mobile phase. The linearity was established in the range 0.1-4% (r > 0.999). The limit of quantitation was 0.1% and the limit of detection was 0.04%. The accuracy of the method was found to be 96.7% (average). For the precision (repeatability), a relative standard deviation value of 2.4% was found. Similarly, a stereoselective HPLC method was also developed and validated for the quantitation of the enantiomer of the starting material used for the synthesis of valsartan, namely (R)-valinebenzyl ester tosylate. Baseline resolution of the enantiomers of valinebenzyl ester tosylate could be achieved on the chiral crown ether column Crownpak CR (Daicel) at 50 degrees C using water-methanol-trifluoroacetic acid (850:150:1, v/v) as a mobile phase. The linearity was established in the range 0.5-5% (r > 0.999). The accuracy of the method was found to be 100.5% (average). For the precision (repeatability), a relative standard deviation value of 3.4% was found. Both methods were found to be suitable for the analysis of the respective analytes.