Abstract
Summary An isocratic RP-HPLC-UV method for analysis of sumatriptan succinate in pharmaceutical dosage forms has been developed and validated. Best separation was achieved on a Thermo Hypersil C4 column (250 mm × 4.6 mm, 5 µm) using a mobile phase of 20 mM potassium dihydrogen phosphate adjusted to pH 4.0 with orthophosphoric acid and acetonitrile (65:35, v/v) at a flow rate of 1.0 mL min−1. UV detection was performed at 227 nm. The method was validated for specificity, linearity, precision, accuracy, limit of quantification, limit of detection, robustness, and solution stability. The calibration plot was linear over the concentration range 25–600 ng mL−1 (r2 = 0.9998) and the limits of detection and quantification were 10 and 25 ng mL−1, respectively. Intra-day and inter-day precision and accuracy were between 1.25 and 2.95% and between −1.15 and 2.47%, respectively. The method was successfully used for analysis of sumatriptan succinate, in the presence of excipients, in orally disintegrating tablets prep...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Acta Chromatographica
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.