Abstract

A simple, specific, sensitive and reproducible high-performance liquid chromatography (HPLC) assay method has been developed and validated for the estimation of odanacatib in rat and human plasma. The bioanalytical procedure involves extraction of odanacatib and itraconazole (internal standard, IS) from a 200 μL plasma aliquot with simple liquid-liquid extraction process. Chromatographic separation was achieved on a Symmetry Shield RP18 using an isocratic mobile phase at a flow rate of 0.7 mL/min. The UV detection wave length was 268 nm. Odanacatib and IS eluted at 5.5 and 8.6 min, respectively with a total run time of 10 min. Method validation was performed as per US Food and Drug Administration guidelines and the results met the acceptance criteria. The calibration curve was linear over a concentration range of 50.9-2037 ng/mL (r(2) = 0.994). The intra- and inter-day precisions were in the range of 2.06-5.11 and 5.84-13.1%, respectively, in rat plasma and 2.38-7.90 and 6.39-10.2%, respectively, in human plasma. The validated HPLC method was successfully applied to a pharmacokinetic study in rats.