Abstract

A reverse-phase high-performance liquid chromatography method was developed to quantify capsaicin (trans-8-methyl-N-vanillyl-6-nonenamid), dihydrocapsaicin (8-methyl-N-vanillylnonanamide) and the main capsaicinoid contents of Capsicum extracts. The chromatographic separation was carried out on a C8 column using isocratic mobile phase consisting of 40% (v/v) acetonitrile and 60% (v/v) orthophosphoric acid solution with flow rate of 1.5 mL/min. The concentration of the eluting compounds was monitored by a diode-array detector at wavelength of 281 nm. The method was evaluated for number of validation characteristics (selectivity, accuracy (confidence intervals <1%), repeatability and intermediate precision (RSD% < 2.5%), limit of detection (LOD), limit of quantification (LOQ) and calibration range). The LOD was 0.25 µg/mL and the LOQ was 0.5 µg/mL. Using methanolic solutions of United States Pharmacopoeia (USP) Capsaicin and Dihydrocapsaicin Reference Standards, the method was linear over the concentration range 0.0005-0.5000 mg/mL for both capsaicinoids. The method was applied to qualify capsaicinoid content of two industrial capsicum extracts according to the USP 29.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call