Abstract

A sensitive and specific liquid chromatography-tandem mass spectrometry assay was developed to quantitate valdecoxib (I) and its hydroxylated metabolite (II) in human plasma. The analytes (I and II) and a structurally analogue internal standard (IS) were extracted on a C 18 solid phase extraction (SPE) cartridge using a Zymark RapidTrace™ automation system. The chromatographic separation was performed on a narrow-bore reverse phase Zorbax XDB-C 8 HPLC column with a mobile phase of acetonitrile:water (50:50, v/v) containing 10 mM ammonium acetate. The analytes were ionized using negative electrospray mass spectrometry, then detected by multiple reaction monitoring (MRM) with a tandem mass spectrometer. The precursor to product ion transitions of m/z 313→118 and m/z 329→196 were used to measure I and II, respectively. The assay exhibited a linear dynamic range of 0.5–200 ng/ml of I and II in human plasma with absolute recoveries from plasma at 91 and 86%, respectively. The lower limit of quantitation was 0.5 ng/ml for I and II. Acceptable precision and accuracy were obtained for concentrations over the calibration curve ranges (0.5–200 ng/ml). Sample analysis time for each injection was 5 min, a throughput of 70 human plasma standards and samples per run was achieved. The assay has been successfully used to analyze human plasma samples to support clinical phase I and II studies.