Abstract

The value of kinematic data for skill assessment is being investigated. This is the first virtual reality simulator developed for liver surgery. This simulator was coded in C++ using PhysX and FleX with a novel cutting algorithm and used a patient data-derived model and two instruments functioning as ultrasonic shears. The simulator was evaluated by nine expert surgeons and nine surgical novices. Each participant performed a simulated metastasectomy after training. Kinematic data were collected for the instrument position. Each participant completed a survey. The expert participants had a mean age of 47 years and 9/9 were certified in surgery. Novices had a mean age of 30 years and 0/9 were certified surgeons. The mean path length (novice 0.76 ± 0.20 m vs. expert 0.46 ± 0.16 m, p = 0.008), movements (138 ± 45 vs. 84 ± 32, p = 0.043) and time (174 ± 44 s vs. 102 ± 42 s, p = 0.004) were significantly different for the two participant groups. There were no significant differences in activating the instrument (107 ± 25 vs. 109 ± 53). Participants considered the simulator realistic (6.5/7) (face validity), appropriate for education (5/7) (content validity) with an effective interface (6/7), consistent motion (5/7) and realistic soft tissue behavior (5/7). This study showed that the simulator differentiates between experts and novices. Simulation may be an effective way to obtain kinematic data.