Abstract
An accurate, precise, simple and selective stability-indicating gradient reverse phase ultra performance liquid chromatographic method has been developed and validated for the quantitative determination of rupatadine and montelukast in pharmaceutical formulation in presence of degradation products. The chromatographic separation was performed on Acquity BEH C8 column (100 mm x 2.1 mm I.D., 1.7 μm) by using mobile phase containing a gradient mixture of solvent A (0.02 M KH2PO4, pH 3.0) and B (90:10 v/v mixture of acetonitrile and water) at flow rate of 0.5 mL/min. The detection was carried out at a wavelength of 240 nm. To establish stability indicating capability of the method, drug product was subjected to the stress conditions of acid, base, oxidative, hydrolytic, thermal and photolytic degradation. The degradation products were well resolved from rupatadine and montelukast. The developed method was validated as per ICH guidelines with respect to specificity, linearity, accuracy, precision and robustness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.