Abstract

An isocratic stability-indicating reversed phase liquid chromatography (RP-HPLC) method for quantitative determination of Zolpidem in the presence of its stressed degradation products in bulk and commercial tablets was developed and validated. Forced degradation studies were carried out on bulk samples and tablet dosage forms of Zolpidem using acid, base, H2O2, heat, and UV light as described by International Conference on Harmonization (ICH) for stress conditions to demonstrate the stability indicating power of the method. A Perfectsil® Target ODS column (3–5 μm, 125 × 4 mm) was used for chromatographic separation. The mobile phase consisted of methanol-50 mM ammonium acetate buffer (pH 3.7, 40: 60, v/v). The flow rate was 0.7 mL/min and UV detection was optimized at 254 nm. The present method linearity for Zolpidem was investigated in the range of 1–20 μg/mL (r = 0.9998). The limits of detection (LOD) and quantification (LOQ) were 400 and 1000 ng/mL respectively. The method selectivity was evaluated by peak purity test using a diode array detector. There was no interference with detection of Zolpidem and its stressed degradation products.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.