A stability indicating HPLC method to determine diltiazem hydrochloride (DTZ) in tablets and compounded capsules was developed and validated according to Brazilian and the International Conference on Harmonization (ICH) guidelines. The separation was carried out on a Purospher Star® C18 (150 x 4.6 mm i.d., 5 µm particle size, Merck Millipore) analytical column. The mobile phase consisted of a 0.05% (v/v) trifluoroacetic acid aqueous solution and a 0.05% trifluoroacetic acid methanolic solution (44:56, v/v). The flow rate was 1.0 mL.min-1 with a run time of 14 minutes. The detection of DTZ and degradation products (DP) was performed at 240 nm, using a diode array detector. The method proved to be linear, precise, accurate, selective, and robust, and was adequate for stability studies and routine quality control analyses of DTZ in tablets and compounded capsules.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.