A simple analytical method for simultaneous determination of phytosterols, cholesterol and squalene in lipid emulsions was developed owing to increased interest in their clinical effects. Method development was based on commonly used stationary (C18 , C8 and phenyl) and mobile phases (mixtures of acetonitrile, methanol and water) under isocratic conditions. Differences in stationary phases resulted in peak overlapping or coelution of different peaks. The best separation of all analyzed compounds was achieved on Zorbax Eclipse XDB C8 (150 × 4.6 mm, 5 μm; Agilent) and ACN-H2 O-MeOH, 80:19.5:0.5 (v/v/v). In order to achieve a shorter time of analysis, the method was further optimized and gradient separation was established. The optimized analytical method was validated and tested for routine use in lipid emulsion analyses.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.