Abstract

In this study, a new, simple, and reproducible high-performance liquid chromatographic method was developed for the determination of pioglitazone in human plasma. After liquid-liquid extraction with diethylether, samples were quantitated on a Nova-Pak C8 column using a mixture of acetonitrile-140mM K2HPO4 (40:60, v/v, pH = 4.45) as mobile phase with UV detection at 269 nm. The flow rate was set at 1.4 mL/min. Ethylparaben was used as internal standard and the total run time of analysis was approximately 7 min. The method was linear over the range of 25-1500 ng/mL of pioglitazone in plasma (r2 > 0.999). The within- and between-day precision values were in the range of 2.4-6.8%. The limit of quantitation of the method was 25 ng/mL. The method was successfully used to study the pharmacokinetics of pioglitazone in healthy volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.