Abstract

GFH009 is a potent, highly selective, small molecule that targets and inhibits the activity of the CDK9/cyclin T1 regulatory complex of P-TEFb. This study aimed to develop and validate a highly selective and sensitive ultrahigh-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method for precise quantification of GFH009 in rat plasma. This method was subsequently employed for conducting toxicokinetic studies of GFH009 in rats. Plasma was prepared using a simple protein precipitation method by acetonitrile. Chromatographic separation of the analytes was achieved on a BEH C18 analytical column with a rapid 3.0 min run time and a flow rate of 0.5ml/min. The calibration curves for plasma samples exhibited excellent linearity over a wide concentration range of 1.0-1,000 ng/ml for GFH009. Intra- and inter-day accuracies were within 92.7-105.7%, and precisions were no more than 6.7%. Furthermore, the analyte demonstrated stability under four different storage conditions, with variations of <15.0%. This study pioneers a methodological innovation by introducing a highly reliable, specific and sensitive analytical method for GFH009 in rat plasma. The successful application of this method in toxicokinetic studies further underscores its significance, offering valuable insights for the methodology of clinical pharmacokinetic research.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call