We developed and validated a sensitive and low sample volume liquid chromatographic-mass spectrometric (LC-MS) method for determination of Promethazine hydrochloride in human plasma (0.5 ml) or urine (0.1 ml). The lower limit of quantification in human plasma and urine was 1.00 ng/ml. The inter- and intra-day precisions (CV %) in both plasma and urine were lower than 10%, the mean method accuracies and recoveries from spiked plasma samples at three concentrations were more than 97%. The developed method was successfully applied to determine Promethazine hydrochloride in human plasma and urine, and proved suitable to clinical pharmacokinetic study.