Abstract

A reversed-phase high-performance liquid chromatographic method has been developed and validated for quantitative determination of deflazacort in tablets and in compounded capsules. Isocratic chromatography was performed on a C18 column with acetonitrile–water 80:20 (v/v) as mobile phase at a flow rate of 1.0 mL min−1. UV detection was at 240 nm. The linearity of the method was good (r > 0.999), as also were intra-day and inter-day precision (RSD 98%). The method was also validated for specificity and robustness. The results showed the proposed method is suitable for its intended use.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.