Abstract

Objective: The aim of this work was to develop and validate a rapid and simple high-performance liquid chromatography method with a diode- array detector (HPLC-DAD) for determination of favipiravir in bulk and tablet formulations.
 
 Methods: The chromatographic analysis was performed at 30 °C with a Poroshell 120EC-C18 column (4.6 x 50 mm, 2.7 µm). The mobile phase was a mixture of 0.1% formic acid in water and 0.1% formic acid in acetonitrile (90:10, v/v). The run time was 5 min at a flow rate of 0.5 mL/min.
 
 Results: The proposed method was successfully validated in terms of precision, accuracy, linearity, robustness, limits of detection (LOD) and quantification (LOQ) parameters. The calibration plot was linear over a concentration range of 10-100 µg/mL. The LOD and LOQ values were found to be 0.58 µg/mL and 2.03 µg/mL, respectively. The average recovery values were found to vary from 99.45 percent to 104.29 percent.
 
 Conclusion: As a result, it was concluded that the developed method can be used successfully in the determination of favipiravir in pharmaceutical preparations.