Abstract

ABSTRACTAbiraterone acetate is a prodrug of Abiraterone widely used for the treatment of metastatic castration resistant prostate cancer. In this study, a simple, sensitive, and rapid stability-indicating reverse phase HPLC method was developed and validated for the determination of Abiraterone acetate in bulk and its pharmaceutical formulation. The method was developed by HPLC using a Hypersil ODS C-18 (150 mm × 4.6 mm, 5 µm) column in a isocratic mode with mobile phase constituted by potassium phosphate buffer and acetonitrile (40:60, v/v%) flow rate was 1.0 mL min−1, column temperature of 30°C, UV detection wavelength 235 nm, and injection volume of 20 µL. The validated parameters were in accordance with FDA and ICH specifications, assay exhibited a linear range of 25–250 µg mL−1 with regression (r2) coefficient 0.9998. The limits of detection and quantification were 0.23 and 0.70 µg mL. Accuracy was between 99.34 and 100.07%. The drug was subjected to various stress conditions like acidic, base hydrolysis, oxidation, thermal, and photolytic degradation. Stress study Abiraterone acetate was found susceptible to degrade under hydrolytic (acid and base) conditions. The proposed method has stability indicating the resolution of the main peak from their degradation peaks. The validated method is suitable for quality control application and reduced analysis time.