Abstract

This study evaluated the safety and efficacy of a novel endovascular thrombectomy device in a swine model of deep vein thrombosis (DVT). The device has an over-the-wire configuration, a manually expandable catching basket, a funnel sheath with a covered stent to minimize the risk of microembolization, and an integrated delivery system. DVT was induced by occluding the right iliac vein with a balloon catheter and injecting thrombin. The novel device was inserted into the inferior vena cava through the right jugular vein access. The device effectively removed the thrombus, restoring venous patency without residual thrombus, vessel injury, or complications. These findings suggest the potential advantages of the novel device over predicate devices. Further clinical evaluation is needed to establish the efficacy of this device in human patients with DVT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.