Abstract

Objective: The aim of this study is to determine the applicability of UV detection method by the sensitive, simply, precise, specific, and low-cost High-Performance Liquid Chromatography (UHPLC) for acetylsalicylic acid in presence of its degradation product. Methods: Method development and determination of dissolution rate for acetylsalicylic acid, were achieved by using reversed-phase liquid chromatography with PDA (Photodiode Array Detector) and TUV (Tunable UV Detector) detection by HCLASS-UHPLC. The new proposed method utilized by the Waters Acquity UHPLC® TUV and PDA systems using a UHPLC column Waters Acquity, BEH, C18, 2.1mm x 50mm, 1.7μm particle size with a mixture of heptane-1-sulphonic acid sodium salt solution which was prepared in acetonitrile: deionized water (15%: 85% v/v) in isocritic mode at a flow rate of 0.5ml/ min, at 25°C with a load of 5μL. The detection for all eluted compounds was carried out at 280nm.