Abstract
A new high-performance liquid chromatography method was developed and validated for the quantitation of gliclazide and repaglinide in pharmaceutical formulations. Determination was performed using a LiChroCART RP-18 column, a mobile phase containing acetonitrile-phosphate buffer (pH 2.1; 60 + 40, v/v), and ultraviolet (UV) detection at 225 nm. Repaglinide was used as an internal standard for gliclazide determination and gliclazide for repaglinide assay. The method was validated with respect to linearity, precision, robustness, ruggedness, accuracy, and specificity. The calibration graphs ranged from 0.015 to 0.09 mg/mL for gliclazide and 0.06 to 0.36 mg/mL for repaglinide. Intra- and interday relative standard deviation values for the standard solutions were 0.70 and 1.01% for gliclazide and 0.78 and 0.93% for repaglinide, respectively. Total recoveries of gliclazide and repaglinide from the laboratory-prepared mixtures were 99.82 +/- 0.58 and 101.50 +/- 0.46% for gliclazide and repaglinide, respectively [mean +/- standard deviation (SD)]. In forced degradation studies, the effect of acid, base, oxidation, UV light, and temperature on both drugs was also investigated. Finally, the method was applied for the quality control of commercial gliclazide and repaglinide tablets. Total recovery was 100.40 +/- 0.35 and 104.46 +/- 0.23% for gliclazide and repaglinide, respectively (mean +/- SD).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.