Abstract
A rapid and simple micellar liquid chromatographic method that does not require use of specific chromatographic columns has been developed and validated for azithromycin determination. The method uses a Hypersil C18 column at 60 °C, 1-butanol–pH 6.86 phosphate buffer solution–water, 15:25:60 (v/v), containing 0.10 M sodium dodecyl sulfate, as mobile phase, and UV-detection at 215 nm. Different characteristics of the method were validated satisfactorily. The specificity, accuracy, linearity, precision (repeatability), and robustness of the method were demonstrated. The method proved suitable for determination of the azithromycin content of capsules and uncoated tablets.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.