Abstract

BackgroundPerforming gait analysis in a clinical setting can often be challenging due to time, cost and the availability of sophisticated three-dimensional (3D) gait analysis systems. This study has developed and tested a portable wireless gait assessment tool (wi-GAT) to address these challenges. AimTo investigate the concurrent validity of the wi-GAT in measuring spatio-temporal gait parameters such as stride length, stride duration, cadence, double support time (DST), stance and swing time compared to a 3D Vicon motion analysis system. MethodsTen healthy volunteers participated in the study (age range 23–30 years). Spatio-temporal gait parameters were recorded simultaneously by the Vicon and the wi-GAT systems as each subject walked at their self-selected speed. ResultsThe stride length and duration, cadence, stance duration and walking speed recorded using the wi-GAT showed strong agreement with those same parameters recorded by the Vicon (ICC of 0.94–0.996). A difference between the systems in registering “toe off” resulted in less agreement (ICC of 0.299–0.847) in gait parameters such as %stance and %swing and DST. Discussion and conclusionThe study demonstrated good concurrent validity for the wi-GAT system. The wi-GAT has the potential to be a useful assessment tool for clinicians.