Abstract
We have developed and validated an accurate, sensitive, and robust LC–MS/MS method that determines the concentration of CRx-102 (the combination of prednisolone and dipyridamole) in human plasma. In this method, prednisolone, dipyridamole, and the combined internal standards (IS) prednisolone-d6 (IS for prednisolone) and dipyridamole-d20 (IS for dipyridamole) were extracted from 100μL human EDTA plasma using methylbutyl ether. Calibration curves were linear over a concentration range of 0.4–200ng/mL for prednisolone and 5–3000ng/mL for dipyridamole. The analytes were quantitatively determined using tandem mass spectrometry operated in positive electrospray ionization in a multiple reaction monitoring (MRM) mode. This validated method has been used successfully in clinical pharmacokinetic studies of CRx-102 in healthy volunteers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.