Abstract

A liquid chromatography-tandem mass spectrometry method for the simultaneous quantification of buprenorphine (BUP), norbuprenorphine (NBUP), buprenorphine glucuronide (BUP-Gluc), and norbuprenorphine glucuronide (NBUP-Gluc) in human urine was developed and fully validated. Extensive endogenous and exogenous interferences were evaluated and limits of quantification were identified empirically. Analytical ranges were 5-1,000 ng/mL for BUP and BUP-Gluc and 25-1,000 ng/mL for NBUP and NBUP-Gluc. Intra-assay and interassay imprecision were less than 17% and recovery was 93-116%. Analytes were stable at room temperature, at 4 degrees C, and for three freeze-thaw cycles. This accurate and precise assay has sufficient sensitivity and specificity for urine analysis of specimens collected from individuals treated with BUP for opioid dependence.

Full Text
Published version (Free)