Abstract

A liquid chromatography–tandem mass spectrometric method (LC–MS/MS) has been developed and validated to determine the concentration of Kendine 91 in mice plasma and tissues. Simvastatin was employed as the internal standard. Separation was performed on a C 8 column, with a mobile phase consisting of methanol and aqueous 10 mM formic acid (73:27 v/v). Both analyte and internal standard were determined using electrospray ionization and the MS data acquisition was via multiple-reaction monitoring (MRM) in positive scanning mode. Quantification was performed using the transitions m/ z 444 → 169 and 441 → 325 for Kendine 91 and simvastatin, respectively. The method was validated with respect to linearity, accuracy, precision, recovery and stability. This assay has been successfully applied to a pharmacokinetic study after intravenous injection of Kendine 91 in mice in a dose of 10 mg/kg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.