Abstract
A simple, rapid, and sensitive high-performance liquid chromatographic method for estimation of efavirenz in human plasma has been developed and validated. Chromatography was performed with C18 analytical column and 50:50 acetonitrile–phosphate buffer (pH 3.5) as mobile phase. Compounds were monitored by UV detection at 247 nm. The retention time for efavirenz was 6.45 min and that for the internal standard, nelfinavir, was 2.042 min. Response was a linear over the concentration range of 0.1 μg–10 μg mL−1 in human plasma. The method was simple, specific, precise and accurate and was useful for bioequivalence and pharmacokinetic studies of efavirenz.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.