Abstract

A dissolution test for telithromycin tablets was validated and developed. In order to choose the most discriminatory one, the conditions to carry out are 900 mL of sodium phosphate buffer at pH 7.5, paddles at 50 rpm stirring speed, time test set to 60 min and using USP apparatus 2 with paddles. The UV spectrophotometric method for determination of telithromycin released was developed and validated. The method presents linearity (r = 1) in the concentration range of 20-60 µg/mL. Precision and recoveries were good, 100.62 and 97.06%, respectively. The method was successfully used for the dissolution test of telithromycin tablets.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.