Abstract

Aim: The 46-gene Prolaris® cell cycle progression test provides information on the risk of prostate cancer progression. Here we developed and validated a 16-gene kit-based version. Methods: RNA was extracted from prostate cancer biopsy tissue. Amplification efficiency, minimum tumor content, repeatability, reproducibility and equivalence with the 46-gene test were evaluated. Results: Amplification efficiencies for all genes were within the acceptable range (90-110%), and samples with ≥50% tumor content were appropriate for the 16-gene test. Results were repeatable (standard deviation: 0.085) and reproducible (standard deviation: 0.115). Instrument, operator and kit lot had minimal impact on results. Cell cycle progression scores from the 46- and 16-gene tests were highly correlated (r=0.969; bias=0.217). Conclusion: The 16-gene test performs consistently and similarly to the 46-gene test.