Abstract

A simple, reproducible and efficient reverse phase high performance liquid chromatographic method was developed for Lisinopril in bulk drug and formulation. A column having 150 × 4.6 mm in isocratic mode with mobile phase containing acetonitrile: phosphate buffer (70:30; adjusted to pH 3.0) was used. The flow rate was 0.8 ml/min and effluent was monitored at 216 nm. The retention time (min) and linearity range (μg/ml) for Lisinopril was (1.510) and (10-35). The developed method was found to be accurate, precise and selective for determination of Lisinopril in bulk and formulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.