Abstract

Nanosized iron oxide particles exhibit osteogenic and radiopaque properties. Thus, iron oxide (Fe3O4) nanoparticles were incorporated into a biodegradable polymer (poly-L-lactic acid, PLLA) to fabricate a composite bone screw. This multifunctional, 3D printable bone screw was detectable on X-ray examination. In this study, mechanical tests including three-point bending and ultimate tensile strength were conducted to evaluate the optimal ratio of iron oxide nanoparticles in the PLLA composite. Both injection molding and 3D printing techniques were used to fabricate the PLLA bone screws with and without the iron oxide nanoparticles. The fabricated screws were implanted into the femoral condyles of New Zealand White rabbits. Bone blocks containing the PLLA screws were resected 2 and 4 weeks after surgery. Histologic examination of the surrounding bone and the radiopacity of the iron-oxide-containing PLLA screws were evaluated. Our results indicated that addition of iron oxide nanoparticles at 30% significantly decreased the ultimate tensile stress properties of the PLLA screws. The screws with 20% iron oxide exhibited strong radiopacity compared to the screws fabricated without the iron oxide nanoparticles. Four weeks after surgery, the average bone volume of the iron oxide PLLA composite screws was significantly greater than that of PLLA screws without iron oxide. These findings suggested that biodegradable and X-ray detectable PLLA bone screws can be produced by incorporation of 20% iron oxide nanoparticles. Furthermore, these screws had significantly greater osteogenic capability than the PLLA screws without iron oxide.

Highlights

  • When the nano-Fe3O4 content was 30% or higher, the material could not be extruded from the extrusion nozzle head of the 3D printing machine

  • The addition of 20% nano-Fe3O4 to the poly-L-lactic acid (PLLA) had slightly reduced the flexural strength of the PLLA polymer

  • Our results showed that the addition of nano-Fe3O4 particles at percentages greater than 20% significantly reduced the ultimate tensile stress of the PLLA

Read more

Summary

Objectives

The purpose of this study was to fabricate biodegradable bone screws using an Fe3O4 nanoparticle/PLLA polymer and test whether or not they exhibited X-ray opacity

Methods
Results
Discussion
Conclusion