Abstract

A novel tissue-equivalent proportional counter (TEPC) based on a gas electron multiplier (GEM) for measuring H*(10) for neutrons was designed and constructed. The pulse height spectra (PHS) of two different neutron sources (a 252Cf source and a AmBe source) were measured using the new TEPC. The measurements were made with the TEPC filled with two different gases (10P gas and a propane-based tissue-equivalent gas) at various pressures. A computer simulation of the new TEPC, based on the Monte Carlo method, was performed to obtain the PHS for the two neutron sources. It is shown that the experimental results agree well with the simulation results for both 252Cf and AmBe neutron sources. Several outstanding problems are discussed and suggestions are made to make the GEM-based TEPC a practical neutron rem meter. The potential advantage of this novel neutron rem meter would be its low weight and compactness.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.