Abstract
Currently, there is alack of methods and tools that efficiently evaluate the auto-feathering junctions created by multileaf collimator (MLCs) for supine volumetric modulated arc therapy (VMAT) craniospinalirradiation (CSI) plans. We have investigated the feasibility of stitching together multi-isocenter fluence maps to then analyze the feathered junctions for patient-specific quality assurance (QA). Furthermore, we investigated the capability of Halcyon for the treatment of CSI patients.Three patients, who previously underwent VMAT CSI treatment on TrueBeam (6-MV flattening filter-free (FFF)) for 36 Gy in 20 fractionswerereplanned for Halcyon. A multi-isocenter approach with only translational superior-inferior shifts was used for both platforms. Each isocenter consists of two full arcs with anterior avoidance sectors, ±5° collimator rotations between arcs, and 5-8 cm of overlapping MLC auto-feathering junctions. All plans were QA'd via electronicportalimaging device (EPID) portal dosimetry and analyzed with a gamma criteria of 3%/3 mm. A variety of plan quality metrics were analyzed to evaluate dose distributions to the target, doses to organs at risk (OARs), and integral dose to the patient. A MATLAB script was developed to stitch the calculated and measured fluence maps in order to perform patient-specific QA for the composite fluence.The Halcyon plans provided highly conformal and homogenous dose distributions to the entire CSI target, superior to the clinical TrueBeam plans, while sparing critical organs with significantly lower values of V10Gy and V18Gy by up to 2% and 2.5%, respectively.Qualitative depictions of vertical dose profiles from the stitched DICOM of the entire CSI target for both planned and delivered fluence maps demonstrated equivalency, with slightly lower average pass rates with Halcyon (97%) compared to TrueBeam(99.9%). This approach to stitch multiple measured versus calculated EPID fluence maps has shown to be a feasible and accurate method and will be helpful for comprehensive VMAT CSI QA on both platforms. Further implementation of this script will be used in examining dosimetric impacts of daily patient positioning errors at MLC auto-feathering junctions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.