Abstract
We developed a novel large-diameter graft "Triplex(®)" that uses a non-biodegradable material as a coating material. This time, in order to demonstrate the physical properties of Triplex(®) grafts, we conducted physical tests in accordance with the international guidelines, using the collagen coated vascular grafts (Hemashield, Boston Scientific, Natick, Massachusetts, USA) as the controls. The grafts were tested with regard to strength (burst strength, circumferential tensile strength, longitudinal tensile strength), suture retention strength, integral water permeability, water leakage (needle puncture, after using clamp), and change in luminal diameter following pacing stress according to ISO7198 and FDA guidance. As indicated by the results, we experimentally demonstrated that uniquely designed vascular graft Triplex(®) led to less blood leakage from the vascular graft and less leakage from the needle puncture, although it has fundamental physical properties comparable to those of the vascular grafts using biodegradable material that has been utilized conventionally in clinical settings. Triplex (®)is expected to play its role as a clinically beneficial next-generation vascular graft.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.