Abstract

Objective To develop a kit for detecting human epidermal growth factor receptor 2 (HER-2) in the human body. Methods The HER-2 kit was evaluated based on an automated magnetic particle chemiluminescence platform. The kit was developed using the double antibody sandwich-complexation method. Results The kit showed a linear range of 0.01–800 ng/mL, with a linear R2 of >0.999. The limit of the blank was 0.0039 ng/mL, and the precision at 1.00 ng/mL was 9.4%. The recovery rate at 10.00 ng/mL was 97.81–101.81%. The negative serum reference range was 0–8.23 ng/mL. Conclusions The kit had a wide linear range, high accuracy, good precision, and high sensitivity, indicating that it has good application prospects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.