Abstract

With the advancement of space exploration in China, mass spectrometers are becoming increasingly important in space missions. A miniature Mattauch-Herzog double focusing mass spectrometer was developed for mass analysis. Based on the study of the fringing field of the analyzer, the ion optical lens software SIMION8.1 was used to simulate the focusing performance of the ion beam. The 90° magnetic analyzer with asymmetric structure is adopted, which has a magnetic flux density of 0.85T. The experiment shows that the ion current is actually affected by the fringing field of electrostatic analyzer which is basically consistent with the simulation results. Besides, the performance indicators of the miniature MS are tested, the mass resolution of N2 is 75.7, the sensitivity of N2 is 4.13 × 105 V/Pa and the minimum detectable partial pressure of N2 is 2.81 × 10−7Pa.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.