Abstract

A drinking water method for seven pesticides and pesticide degradates is presented that addresses the occurrence monitoring needs of the US Environmental Protection Agency (EPA) for a future Unregulated Contaminant Monitoring Regulation (UCMR). The method employs online solid phase extraction-liquid chromatography-tandem mass spectrometry (SPE-LC-MS-MS). Online SPE-LC-MS-MS has the potential to offer cost-effective, faster, more sensitive and more rugged methods than the traditional offline SPE approach due to complete automation of the SPE process, as well as seamless integration with the LC-MS-MS system. The method uses 2-chloroacetamide, ascorbic acid and Trizma to preserve the drinking water samples for up to 28 days. The mean recoveries in drinking water (from a surface water source) fortified with method analytes are 87.1-112% with relative standard deviations of <14%. Single laboratory lowest concentration minimum reporting levels of 0.27-1.7 ng/L are demonstrated with this methodology. Multi-laboratory data are presented that demonstrate method ruggedness and transferability. The final method meets all of the EPA's UCMR survey requirements for sample collection and storage, precision, accuracy, and sensitivity.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call