Abstract

An imaging system of thermal neutrons for an investigation of digital neutron radiography has been developed and tested. Some characteristics obtained on a neutron radiography beam of an experimental reactor are reported. The coordinates of each event are determined in this system. After processing in a LSI 11 23 computer, a radiograph, accumulated in a histogramming memory of 64 K 16-bit words, is presented on a colour display. A 230 × 180 mm 2 low-pressure multistep chamber is used as a detector. Neutron conversion takes place in a 6 μm boron layer enriched to 86% in 10B. The detection efficiency of thermal neutrons is no less than 3%. The count rate of the system reaches up to 2 × 10 5 events per second. A radiograph can be obtained within 10 minutes. The sensitivity of this system to gamma-background is low. One event/s is detected for a background of 1 R/h. The spatial resolution is found to be 0.7 mm (FWHM) using a cadmium knife edge. The integral nonlinearity is less than 0.4%. The possibility of using a hydrogeneous converter in this system for neutron radiography is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.