Abstract
The performance of a uric acid determination kit has been evaluated for five months, under routine conditions, in a General Hospital Biochemical Laboratory. An anomalous increment in fresh serum blanks was noted in the kit when first introduced. This interference, probably due to alcohol dehydrogenase contamination, was corrected by addition of 50 mmol/l (NH4)2SO4 to the reagent. Results obtained with this modified reagent correlate perfectly with those obtained with modified kits by Smith Kline Instruments (SKI), and with many other determination methods. Correlations are discussed and explanations for differences in statistical data are offered. Within run and between run precision data are presented. The kit fits perfectly with the needs of centrifugal fast analyzers and discrete micro analyzers, on account of its speed, reliability and precision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.