Abstract

We introduced and validated a method to encase guiding optical coherence tomography (OCT) probes into clinically relevant 36G polyimide subretinal injection (SI) cannulas. Modified SI cannulas presented consistent flow capacity and tolerated the typical mechanical stress encountered in clinical use without significant loss of sensitivity. We also developed an approach that uses a micromanipulator, modified SI cannulas, and an intuitive graphical user interface to enable precise SI. We tested the system using ex-vivo porcine eyes and we found a high SI success ratio 95.0% (95% CI: 83.1-99.4). We also found that 75% of the injected volume ends up at the subretinal space. Finally, we showed that this approach can be applied to transform commercial 40G SI cannulas to guided cannulas. The modified cannulas and guiding approach can enable precise and reproducible SI of novel gene and cell therapies targeting retinal diseases.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.