BackgroundA large variety of electronic hand hygiene monitoring systems (EHHMS) are being developed and applied in health care settings. Monitoring hand hygiene (HH) opportunity at bed-level has been the key technical challenge. Accuracy evaluation needs more attention as the prerequisite upon widespread acceptance and adoption. MethodsFor the first time, we explored, debugged and upgraded an EHHMS based on ultra-wide bandwidth (UWB) which can obtain HH opportunities at bed-level. The real-time positioning and electronic fence of UWB technology was applied for EHHMS. The accuracy of EHHMS was compared with the simultaneous manual direct observations in real-world clinical setting. Sensitivity and specificity were calculated for EHHMS capturing HH action and opportunity. ResultsTwo generations of EHHMS were constructed. For the first generation, the system properly recorded 84% and 78% of the pre-identified HH actions and opportunities performed by experimenters. For the second generation, sensitivity and specificity of the system capturing HH action were 89% (84.83-92.36) and 100% (98.26-100.00), respectively. For capturing HH opportunity, the system showed the sensitivity and specificity of 86.52% (82.52-89.89) and 88.10% (84.14-91.36)), respectively. ConclusionThe EHHMS based on UWB could accurately identify HH action and opportunity with equivalent accuracy compared with simultaneous direct observation.