Abstract
Abstract Modification of a commercially available TG/DTA instrument enabled development of a working TG/DTA/Raman system. The TG/DTA was modified by replacing a portion of the furnace wall above the sample and reference holders with a double layer insulated quartz window. The quartz window provided ready access for Raman spectroscopy, enabling simultaneous collection of TGA, DTA, and Raman data during thermal analytical experiments. Insertion of the quartz glass window did not impact TGA or DTA performance as demonstrated by the comparison of baseline drift and temperature calibration for modified and unmodified furnaces.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.