Abstract

A novel dissolution apparatus was developed for medicated chewing gum products. A prototype gum product containing phenylpropanolamine hydrochloride (PPA) was used to evaluate the apparatus. The apparatus consists of a conical Teflon base and a rotating, ribbed Teflon plunger suspended in a dissolution vessel. Parameters evaluated were rotation speed, plunger frequency, medium volume, medium type, medium sampling location, number of plunger ribs, and number of gum pieces. Samples were taken over a 20-min period and samples were analyzed by HPLC. Cumulative percentage released-versus-time profiles were obtained for each parameter evaluated. Statistical analysis of the gum product indicated that the only significant differences occurred at the lowest rotation speed and lowest plunger frequencies. A Level A correlation was found between the in vitro release profile for the 20-rpm and 30-cycles/min plunger frequency and the in vivo chew-out study.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call