Abstract
We have developed a competitive assay to measure the estrogen-regulated protein pS2. A monoclonal pS2 antibody (mAb) and a biotinylated pS2 peptide are used, with time-resolved fluorometry as a detection technique. The assay has a detection limit of 16 ng/mL and is precise (within-run and day-to-day Cvs 3-12%). We used this assay to determine steroid hormone activity of six steroids in cell culture, both in terms of time course and dose response. pS2 concentrations in the tissue culture supernatant of the BT-474 breast carcinoma cell line were significantly higher when estradiol was the stimulating steroid. There was a significant time course and dose response observed for estradiol, but not for the other steroids. The availability of a sensitive, reliable, and convenient method for quantifying pS2 will allow for many research applications including the screening of natural and synthetic compounds for putative estrogenic activity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.