Abstract
This study presents the development and the comparison of high accuracy methods for uranium isotope determination by thermal ionization mass spectrometry. Two methods for uranium minor isotope ratio determination were compared in term of accuracy, analysable quantity, analysis time and versatility: the total evaporation and the classical method with multi-dynamic sequences. The mathematical correction of the abundance sensitivity and the detector calibration within the classical method helps decreasing the uncertainties and the biases compared to the total evaporation method. This comparative study was conducted within the framework of the “2017 Nuclear Material Round Robin” participation organized by the International Atomic Energy Agency.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.