Abstract

Cadmium zinc telluride (CZT) detectors are suitable for various applications due to the good energy resolution and the simple pixilation to achieve high spatial resolution. Our group is developing a two-panel head and neck dedicated positron emission tomography system based on CZT detectors. Each panel will consist of 150 CZT crystals (4×4×0.5 cm3) covering an area of 20×15 cm2 in an edge-on configuration to achieve high detector efficiency at 511 keV. In this work, we present the design and development of a full data acquisition chain that enables a low noise and compact readout for each panel. The initial results of the readout circuit were quantified using a 1 kHz square wave test pulse. The pulse amplitude was chosen to generate approximately the same amount of charges as a 511 keV photon would provide in CZT. The best-case FWHM electronic noise at 511 keV was measured to be 0.69% ± 0.16% (3.52 ± 0.81 in keV units after conversion). The FWHM electronic noise at 511 keV for a complete DAQ chain was 4.33% ± 0.30% (22.13 ± 1.53 in keV units).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.